可昌新技 台灣布魯克 分光儀 | |
公司簡介 產品介紹 技術服務 聯繫我們 | |
布魯克 氣體分析儀 :: G4 Phoenix 產品特性 規格參數 常見問題 |
|
High Performance Analyzer for diffusible HydrogenMeasuring range0.05 – 1000 μg/g hydrogen Hot extraction furnaces Infrared furnace:900°C max., quartz tube diam. 30 mm Wire-heated furnace: 1100°C max., quartz tube diam. 18 mm Analysis time* approx. 3 min – 2 h For weld samples typically 15 – 20 min * Depending on sample material, weight and analysis temp. Resolution 0.01 μg/g Precision* ±0.05 μg/g or 1 % RSD (whichever is greater) * Depending on sample material, weight and analysis temp. Carrier gas Nitrogen 99,999% purity, min. pressure 2 bar (~50 psi) Cooling water approx. 1 l/min (only for IR furnace and only when heating) Dimensions & weight 630 x 640 x 650 mm (WxDxH); 50 kg 24.8 x 25.2 x 25.6 inches; 110 lbs. Electrical supply Analyzer: 230 VAC, 50/60 Hz, 1 kVA Peripherals: 230 VAC, 50/60 Hz, 300 VA PC requirements Standard WindowsTM-PC, operating system Windows XP/7* HDD > 200 GB, DVD drive, 2 serial ports RS 232 *Windows is a registered trademark of Microsoft Corporation Optional mass spectrometerDetection limit50 – 100 ppb hydrogen at 1 g sample weight m/z range 1 - 100 amu, optimized for mass 2 (hydrogen) Calibration With H2 gas via the gas dosing unit G4 PHOENIX DH Electrical supply 230VAC, 50/60 Hz, 250 VA Dimensions and weight 630 x 640 x 480 mm (WxDxH), approx. 60 kg 24.8 x 25.2 x 18.9 inches; 132 lbs. |
|