High Performance Oxygen, Nitrogen and Hydrogen AnalyzerMeasuring ranges*Oxygen: 0.1 – 250 μg/g and optional 0.02 – 0.5 % Nitrogen: 0.1 μg/g - 0.5 % Hydrogen: 0.01 - 1000 μg/g * Can be extended by adjusting the sample mass Nominal sample weight 0.5 g Analysis time* 50 s nominal For diffusible hydrogen up to 60 mins * Depending on sample material, weight and analysis temp. Resolution 0.01 μg/g Precision* NDIR (CO): ±0.5 μg/g or 1 % RSD** TCD (N2, H2): ±0.05 μg/g or 0.5 % RSD** * Depending on sample weight and concentration ** Whichever is greater Carrier gas Nitrogen or Argon 99,999% Helium 99.996% 2 bar (~50 psi) pressure Compressed air 5 bar (~72.5 psi) min. pressure oil and water free Water cooling approx. 4 l/min (only when heating) (for IR furnace 1 l/min) Dimensions & weight 800 x 600 x 830mm (wxdxh); approx. 120 kg 31.5 x 23.6 x 32.6 inches; 264.5 lbs. Power supply Analyzer: 400 VAC, 50/60 Hz, 7 kVA Peripherals: 230 VAC, 50/60 Hz, 800 VA PC requirements Standard WindowsTM-PC, operating system Windows XP/7* HDD > 200 GB, DVD drive, 2 serial ports RS 232 *Windows is a registered trademark of Microsoft Corporation Optional mass spectrometer for hydrogenDetection limit50 – 100 ppb hydrogen at 1 g sample weight With higher sample mass, e.g. 10 g, the detection limit is 10 times lower m/z range 1 - 100 amu, optimized for mass 2 (hydrogen) Calibration With H2 gas via the gas dosing unit G8 GALILEO Electrical supply 230VAC, 50/60 Hz, 250 VA Dimensions and weight 630 x 640 x 480 mm (WxDxH), approx. 60 kg 24.8 x 25.2 x 18.9 inches; 132 lbs. |
|